7.19 Inside the glove box:
Then check
that the cell cover is firmly seated on the
housing and that no gap is visible.
7.20 Inside the glove box:
Now screw in
the spring load into the cell base.
7.21 Inside the glove box:
Pick up some
lithium with the reference pin. Make sure
that the hole of the reference pin is
completely filled with lithium metal.
Incomplete filling may result in scatter/
noise of the WE potential.
Make sure that the reference pin and the
PTFE ferrule are not corroded or damaged.
The PTFE ferrule must be white and must
not show any black coloration. Replace if
necessary.