7.10 Inside the glove box:
: Put this
assembly into the cell base body. Make
sure that the two grooves at the frit flange
and the cell base body are properly
aligned.
Note:
Make sure that the frit flange is fully
inserted. You may otherwise damage the
cell.
7.11 Inside the glove box:
: Add 320-420
μl of electrolyte using an pipette.
7.12 Inside the glove box:
: Place the
working electrode with the active side
pointing downwards on top of the T-Frit.