User Manual
ECD-3-nano
Release 1.52
6.8
When dissassembling the dilatometer for
the first time, remove the stiff plate below the
cover flange. This plate is for transport only.
For the actual experiment, replace the plate by
the provided metal membrane.
6.10
Now the frit flange with the PTFE-
Seal and the piston in the middle are
visible.
6.9
Remove the stiff plate or membrane
from the cell body
6.11
Pull the frit flange out of the cell body.