EMC Information
SC430-PTR has the capabilities of capturing images with the camera and outputting the images.
Intended Use Environments
The SC430-PTR is intended to be used in professional healthcare facility environments such as
clinics and hospitals (including use in the vicinity of high-frequency surgical equipment such as
electrosurgical knives).
The following environments are not suitable for using the SC430-PTR.
•
Home healthcare environments
•
In the vicinity of short-wave therapy equipment
•
RF shielded room of MRI medical equipment systems
• In shielded special environments
• Installed in vehicles including ambulances
• Other special environments
WARNING
The SC430-PTR requires special precautions regarding EMC and during installation. You need
to carefully read the EMC Information and the “PRECAUTIONS” section of this document, and
observe the following instructions when installing and operating the product.
The SC430-PTR should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the equipment or system should be observed to check for normal
operation in the con guration in which it will be used.
When using portable RF communication equipment, keep it 30 cm (12 inches) or more away from
any part, including cables, of the SC430-PTR. Otherwise, degradation of the performance of this
equipment could result.
Anyone who connects additional equipment to the signal input parts or signal output parts when
con guring a medical system is responsible for ensuring that the system complies with the
requirements of IEC/EN60601-1-2.
Images may be distorted if the product is used near a device such as high-frequency surgical
equipment. Check in advance to ensure that no problems occur during use.
Be sure to use cables that satisfy the following requirements.
Use of cables that do not satisfy the requirements could result in increased electromagnetic
emissions, reduced electromagnetic immunity of this equipment, and incorrect operation.
Cables
Max. Cable Length
Shielding
DC Cord
2.5 m
Shielded
AC Cord
2 m
Unshielded
BNC Cable (SDI)
30 m
Shielded
HDMI Cable
5 m
Shielded
RS-232C Cable
5 m
Shielded
USB Cable
5 m
Shielded
Camera Cable
25 m
Shielded
LAN Cable
30 m
Unshielded
Audio Cable
5 m
Shielded
En
gli
sh
49
Appendix
Содержание CuratOR SC430-PTR
Страница 59: ...English 59...