- 50 -
Chapter 5
Troubleshooting
Error Code
Description
0050
•
Failed to initialize the built-in calibration sensor. External light may have entered the built-in calibration sensor. Ensure that light does not enter
the sensor, and then execute SelfCalibration again.
•
Communication with the built-in calibration sensor may have failed. Turn off the main power, wait for a few minutes before turning the power
back on, and then execute SelfCalibration again.
0060
0061
•
The built-in calibration sensor may not have worked correctly.
•
Check whether there is any foreign object near the built-in calibration sensor.
•
Execute SelfCalibration again.