Use
Before initial use, brand-new tools and accessories must under go the entire reprocessing process and the batteries should
be charged.
Inserting the Battery in the Battery Casing
To ensure sterility, the battery is inserted into the battery casing by two people, one of whom is wearing sterile garments.
1. The person with the sterile garments holds the sterile battery casing. If the casing is not opened, the same person
presses the central button to unlock
(Fig. 3) and the housing cover with slide (Fig. 4) in arrow direction
(Fig. 5)
and opens the lid. In doing so, fold the lid outwards with the locking mechanism.
Figure 2
Figure 4
Figure 5
Figure 3
2. The person wearing the sterile garments places the sterile funnel (item no. 195085) on the battery casing (Fig. 6) and
checks if it is seated correctly (Fig. 7). The sterile cover ensures that the unsterile battery does not contact the outside of
the sterile casing.
Figure 6
Figure 7
3.
The person not wearing sterile garments carefully guides the unsterile battery through the sterile funnel (item no.
195085), (Fig. 8).
The orientation of the battery is correct when the battery with groove slots and arrow symbol is inserted into the
corresponding spring profi le receptacle in the handpiece (Fig. 7).
The same person presses it completely into the battery casing to ensure a correct seat (Fig. 9). This person may not contact
the outside of the battery casing.
Figure 9