24
USER MANUAL
ORTHOVET PLUS BATTERY DRILL SYSTEM
T +49 7461 96 580 0 | F + 49 7461 96 580 90 | [email protected] | www.eickemeyer.com
5. TROUBLESHOOTING
Problem
Possible causes
Solution
Handstück does not start up.
Battery is dead.
Charge the battery or replace it with a
charged battery.
The tool was not cooled of after
sterilization.
Let the tool cool to room temperature.
Mode selector switch is on „OFF“.
Turn the mode selector on “R” or “L”
position.
No contact between the handpiece and
the battery.
Charge the battery or replace it with a
charged battery.
Handpiece does not have enough
power.
Battery is dead.
Charge the battery or replace it with a
charged battery.
Handpiece stopps suddenly.
The handpiece has overheated
(overheating protection is activated).
Wait until the handpiece has cooled
down.
Battery is dead.
Charge the battery or replace it with a
charged battery.
Attachments / Chucks cannot be
coupled to to the unit.
The attachment coupling is blocked by
deposits.
Remove solid objects with a pair
tweezers.
Precaution
: When removing objects,
turn the mode selector switch to „OFF“.
Toll (saw blade, drill, burr etc.) cannot
be coupled or only with diffi culty.
Shaft geometry of the attachment or
tool is damaged.
Replace the attachment or tool or send
it to your EICKEMEYER® service offi ce.
Oscillating Saw-Chuck vibrates too
much.
The saw blade locking Mechanism is
not tight.
Tighten the locking knob of the Saw
blade coupling.
The KIRSCHNER-Wire is inserted in the
handpiece and cannot be moved
forward.
The KIRSCHNER-Wire was inserted
from the rear.
Lock the handpiece by turning the
mode selector switch to “OFF” remove
the attachment, hold the drive shaft
opening down and shake out the
KIRSCHNER-Wire.
Bone and tool heat up during surgery.
Cutting edges of the tool are blunt.
Replace the tool.
It is diffi cult to close the battery casing
The battery casing seal has become dry
from repeated.
Lubrifi cate the seal as described on
page 19.
The trigger is diffi cult to move.
The trigger shaft needs to be
lubrifi cated.
Lubrifi cate the trigger as described on
page 19.
If the recommended solutions do not work, send the power tool back to EICKEMEYER® service center.
For further technical questions or information on our services, please contact your EICKEMEYER® representative.