Change Saw Blade
1. Lock the machine with the Rotating Ring at the handpiece put in position “OFF”.
2. Press down the locking knob continuously.
3. Lift and remove the saw blade.
4. Use a slight amount of pressure to insert the new saw blade and turn it to one of the 5 desired position.
5. Place your thumb on the saw blade coupling to hold the saw blade. Insert the saw blade in such a way that the perforated
holes in the saw blade fi t in the pins of the holder. Release the locking knob.
6. Unlock the power tool.
Positioning the Saw Blade
The saw blade can be adjusted in the desired position vertically and horizontally at an angle of 45° (see the earlier sections
“Positioning the saw attachment” and “Changing the saw blade” (Fig. 21).
Figure 21
Using the Oscillating Saw Attachment
The saw blade must already be oscillating when the saw is applied to the bone. Do not apply strong pressure to the saw blade
as this will delay the cutting process and the saw teeth will catch in the bone. Optimal saw performance is achieved by
moving the power tool slightly back and forth in the plane of the saw blade so that the blade oscillates beyond the bone on
both sides. Very precise cuts can be made when the saw blade is guided steadily. Imprecise cuts arise due to used blades,