ORTHOVET PLUS BATTERY DRILL SYSTEM
T +49 7461 96 580 0 | F + 49 7461 96 580 90 | [email protected] | www.eickemeyer.com
Function Control
• Visually inspect for damage and wear.
• Check the handpiece controls for smooth operation and function.
• All movable parts should be moving smoothly. Check that the trigger does not remain blocked in the handpiece when
pressing on them. Check that no residuals prevent the movable parts from moving smoothly.
• Check the coupling sleeves of the handpiece and attachments for smooth operation and check for function together with
instruments such as cutting tools.
• Check instruments for correct adjustment and functioning prior to every use.
Packaging, Sterilization and Storage
Packaging
Put cleaned, dry products into the proper places in the EICKEMEYER® tray (item no. 195086) and put the tray into the
container (item no. 185555). Care should be taken in order to protect implants as well as pointed or sharp instruments from
contact with other objects that may damage the surface or the Sterile Barrier System.
Sterilization
Note
: EICKEMEYER® recommends the use of the tray (item no. 195086) specially designed for the system.
The OrthoVet Plus System may be resterilized using validated steam sterilization methods (ISO 17665 or national standards).
EICKEMEYER’s® recommendations for packed devices and cases are as follows.
Cycle type
Sterilization exposure
time
Sterilization exposure
temperature
Drying time
Saturated steam-forced
air removal (pre-vaccum)
(minimum 3 pulses)
Minimum 4 minutes
Minimum 132 °C
Maximum 138 °C
20 – 60 minutes
Minimum 3 minutes
Minimum 134 °C
Maximum 138 °C
20 – 60 minutes
Drying times generally range from 20 to 60 minutes due to differences in packaging materials (Sterile Barrier System, e. g.,
wraps or reusable rigid container systems), steam quality, device materials, total mass, sterilizer performance and varying cool
down time.
Precautions
•
The following maximum values may not be exceeded: 138 °C.
Higher values can damage the sterilized products.
• Do not accelerate the cooling process.
• Hot air, ethylene oxide, plasma and formaldehyde sterilization are not recommended.