66
GPS 3 User Manual | 8103002-56
•
2018-07-17
2.
With GPS Solver, solve the file as you would for a
patient by identifying, verifying, and adjusting
targets instead of sensors.
3.
If the Sensor Errors are under 1.5 mm, your
calibration is good. If you have questionable
Sensor Errors, use your judgment to recalibrate the
system, or not. If unsure, contact EGI Technical
Support (Table P-1).
•
The sensor should now turn green. If it doesn’t, contact EGI
Technical Support (Table P-1).
2
After all sensors have turned green, rotate through the 3D
model view to make sure nothing was missed.
3
Verify in the Status pane that all 11 camera views show as
verified and all Net sensors show as adjusted.
5.2.5
Specify the Fiducials (Nasion, LPA, and RPA)
You will need to mark each fiducial point in two camera views.
1
Use the 3D model view and the Status pane to look for the
best camera view of each fiducial point (nasion, right
preauricular, and left preauricular).
2
Carefully locate a fiducial in the best camera view, right click
on the center dot and select the correct fiducial label (Nasion,
LPA, or RPA).
Содержание Geodesic Photogrammetry System
Страница 1: ......
Страница 10: ...10 GPS 3 User Manual 8103002 56 2018 07 17...
Страница 24: ...24 GPS 3 User Manual 8103002 56 2018 07 17...
Страница 44: ...44 GPS 3 User Manual 8103002 56 2018 07 17...
Страница 50: ...50 GPS 3 User Manual 8103002 56 2018 07 17 Figure 5 1 GPS Acquisition s Capture Images window...
Страница 55: ...GPS 3 User Manual 8103002 56 2018 07 17 55...
Страница 71: ...GPS 3 User Manual 8103002 56 2018 07 17 71...
Страница 73: ...GPS 3 User Manual 8103002 56 2018 07 17 73 Figure A 2 64 channel GSN sensor layout...
Страница 74: ...74 GPS 3 User Manual 8103002 56 2018 07 17 Figure A 3 128 channel GSN sensor layout...
Страница 75: ...GPS 3 User Manual 8103002 56 2018 07 17 75 Figure A 4 256 channel GSN sensor layout...
Страница 97: ...GPS 3 User Manual 8103002 56 2018 07 17 97...
Страница 98: ...98 GPS 3 User Manual 8103002 56 2018 07 17...