60
4.7.3 IR
Select IR for the test type and use the enter key (
↵
)
to save the Test Type and move to the next
test parameter:
Note: Charge-Lo: The parameter can be set automatically or manually. To perform an auto Charge-
Lo connect the DUT and all the test leads as needed to perform a real test. Press the test button
on the front panel. Reading will be displayed on the screen when the instrument is performing the
auto Charge-Lo. Once the instrument is has finished performing the auto Charge-Lo the numeric
value will be displayed on the screen and will be automatically saved to be used when the actual
test is performed.
4.7.4 GND
Select GND for the test type and use the enter key (
↵
)
to save the Test Type and move to the next
test parameter: