97
Calibration of AC Hipot Voltage
1.
Connect the standard 5KVAC kilovolt meter from H.V. to Return.
2.
When the standard voltmeter is connected, press TEST to start the calibration process.
3.
Enter Standard Voltage Reading.
Calibration of DC Hipot Voltage
1.
Connect the standard 6KVDC kilovolt meter from H.V. to Return.
2.
When the standard voltmeter is connected, press TEST to start the calibration process.
3.
Enter Standard Voltage Reading.
Calibration of 30mA AC Total Current Range
1.
Connect the 100K
load in series with the standard current meter.
2.
When the load is connected, press TEST to start the calibration process.
3.
Enter Standard Current Reading.
Calibration of 3.5mA AC Total Current Range
1.
Connect the 100K
load in series with the standard current meter.
2.
When the load is connected, press TEST to start the calibration process.
3.
Enter Standard Current Reading.
Calibration of 9.99mA DC Current Range
1.
Connect the 100K
load in series with the standard current meter.
2.
When the load is connected, press TEST to start the calibration process.
3.
Enter Standard Current Reading.
Calibration of 3.500mA DC Current Range
1.
Connect the 100K
load in series with the standard current meter.
2.
When the load is connected, press TEST to start the calibration process.
3.
Enter Standard Current Reading.
Calibration of 350uA DC Current Range
1.
Connect the 1M
load in series with the standard current meter.
2.
When the load is connected, press TEST to start the calibration process.
3.
Enter Standard Current Reading.
Calibration of 35uA DC Current Range
1.
Connect the 50M
load in series with the standard current meter.
2.
When the load is connected, press TEST to start the calibration process.
3.
Enter Standard Current Reading.
Calibration of 3.5uA DC Current Range
1.
Connect the 100M
load in series with the standard current meter.