4.2.4 Memory, Step, and Default System Parameter Restore
Restoring the instrument’s System parameters will overwrite all
memories and steps with ACW default parameters!
Press the two bottom soft keys and power the instrument at the same time, then press
the soft key next to the word “Yes”. All memories and steps will be loaded with the
ACW default parameters and the system parameters will be set to the factory defaults.
The following System parameters will not be affected by the system initialization:
Setup System
Cal Alert
No change
Cal Date
No change
Cal Due
No change
Alert
No change
Date m/d/y
No change
Time
No change
4.3 Test Setup
1. From the
Perform Tests screen, press the “Menu” soft key. The Main Menu
screen will now be displayed.
2. From the Main Menu screen,
press the “Menu” soft key. Results, Test and
System selections will now be displayed.
3. From this screen, pres
s the “Test” soft key. The Test Parameter Review screen
will now be displayed.
4. From this screen all relevant test parameters may be edited and saved. Refer to
section
4.3.4. Default Test Parameters
for a complete list of test parameters.
4.3.1 Test Setup Soft Keys
Directional soft keys
,
,
The “
,
,
” soft keys are used to scroll the cursor to the different system parameters.
+/- keys
The “+” and “-“ soft keys are used to increase or decrease numerical values or toggle
settings ON and OFF.
Enter key
The “Enter” soft key is used to store a specific value for each test parameter.
Edit key
The “Edit” key is used to access a specific parameter and allow it to be changed.
Esc key
The “Esc” key is used to return to the Test Parameter Review screen during
parameter editing without saving the new parameter value.
WARNING