D0300040001EN - Computer interface
Page 46
SSC (set sensor control)
Setting the type of sensor control for CPG.
Example:
SSC[a, b, c, x.xxxxE±xx, x.xxxxE±xx]<CR>
Response:
OK<CR>
c
Sensor switch-off type channel
0 = Manual
1 = External
2 = Self-monitoring
3 = Through channel 1
4 = Through channel 2
x.xxxxE±xx
Switch-on value in the current unit of measurement
x.xxxxE±xx
Switch-off value in the current unit of measurement
Parameter
Description
a
Channel number
3 = Channel 3 (CPG)
b
Sensor switch-on type channel
0 = Manual
1 = External
2 = n.a.
3 = Through channel 1 (PRG 1)
4 = Through channel 2 (PRG 2)
c
Sensor switch-off type channel
0 = Manual
1 = External
2 = Self-monitoring
3 = Through channel 1 (PRG 1)
4 = Through channel 2 (PRG 2)
x.xxxxE±xx
Switch-on value in the current unit of measurement
x.xxxxE±xx
Switch-off value in the current unit of measurement
Parameter
Description
Содержание PGC201
Страница 10: ...This page has been intentionally left blank D0300040001EN Contents Page vi ...
Страница 14: ...D0300040001EN Safety Page 4 This page has been intentionally left blank ...
Страница 30: ...D0300040001EN Installation Page 20 This page has been intentionally left blank ...
Страница 38: ...D0300040001EN Operation Page 28 This page has been intentionally left blank ...
Страница 46: ...D0300040001EN Parameters Page 36 This page has been intentionally left blank ...
Страница 58: ...D0300040001EN Computer interface Page 48 This page has been intentionally left blank ...
Страница 62: ...D0300040001EN Storing and waste disposal Page 52 This page has been intentionally left blank ...
Страница 63: ...This page has been intentionally left blank ...
Страница 64: ...This page has been intentionally left blank ...
Страница 67: ...This page has been intentionally left blank ...
Страница 68: ...P800 80 000 Issue T edwardsvacuum com ...