In order to ensure a stable but also correct adjustment of the 100% point, the
adjustment potentiometer (100%) should be turned further by 90° in the clockwise
direction after the indication of 1000 mbar has been displayed. Here the over-range
indicator should just come on.
Evacuate the vacuum system to a pressure < 5 x 10
-4
mbar and adjust the "0"
potentiometer on the PRG gauge head so that the following reading is obtained:
Figure 12 Zero adjustment for the PRG gauge head
Vent the vacuum system and check the 100% setting once more. Correct any deviation
with the 100% potentiometer.
If a correction of the 100% setting was necessary, then repeat the Zero adjustment.
Parameter mode
Selection
By pressing the (PARA) pushbutton for approximately 2 seconds the instrument changes from
the measurement mode to the parameter mode. The Para indicator comes on for the channel
selected in each case. When the instrument is running in the parameter mode and if no
pushbutton is operated for 10 seconds, then the instrument will automatically return back to
the measurement mode. The Para indicator for the channel which was selected in each case
is turned off.
Parameter groups
In the parameter mode you have access to different parameters. You may view these
parameters or change these with the aid of the arrow pushbuttons. In this way you can
configure the instrument.
shows all available parameters.
Adjustment too high
Adjustment is correct
Adjustment is too low
Содержание PGC201
Страница 10: ...This page has been intentionally left blank D0300040001EN Contents Page vi ...
Страница 14: ...D0300040001EN Safety Page 4 This page has been intentionally left blank ...
Страница 30: ...D0300040001EN Installation Page 20 This page has been intentionally left blank ...
Страница 38: ...D0300040001EN Operation Page 28 This page has been intentionally left blank ...
Страница 46: ...D0300040001EN Parameters Page 36 This page has been intentionally left blank ...
Страница 58: ...D0300040001EN Computer interface Page 48 This page has been intentionally left blank ...
Страница 62: ...D0300040001EN Storing and waste disposal Page 52 This page has been intentionally left blank ...
Страница 63: ...This page has been intentionally left blank ...
Страница 64: ...This page has been intentionally left blank ...
Страница 67: ...This page has been intentionally left blank ...
Страница 68: ...P800 80 000 Issue T edwardsvacuum com ...