Technical data
B722-40-880 Issue H
2.3
Vent gas specification and vent control data
Although the pump may be vented to atmosphere, high relative humidity of the air may greatly increase the
subsequent pump-down time. To reduce pump-down times vent with dry, clean gases. Refer to
for a
description of the vent options and the vent valve connection and refer to
for configuring the venting
options.
Figure 2 - Maximum allowed rate of pressure rise during venting: pressure against time (with pump initially at
full rotational speed)
Table 5 - Vent gas specification and vent control
Vent gas specification and control
Reference data
Vent gas
Dry air, nitrogen, argon or other inert gases
Maximum dew point at atmospheric pressure
-22 °C
Maximum size of particulates
1
µ
m
Maximum concentration of oil
0.1 parts per million
Recommended time for rotational speed to reach 50%
> 15 seconds
Maximum allowed rate of pressure rise
Refer to