Installation
3.13.3
Vent options
Note:
The Axial Emergency condition is only available on the EXC300M Controller. On other Controllers, the
position of DIP switch 2 can be ignored.
DIP switches 2 to 4 are used to select the vent-valve control options (refer to
). Select the required vent
options as shown in
Note however, that if the electrical supply to the Controller fails, the Controller will always switch off the electrical
supply to the vent-valve when the EXT pump speed falls to below 50% of full speed (see
).
Table 10 - Selection of vent-valve control options
DIP switch positions
Vent option(s) selected
2
(Vent on
Axial
Emergency)
3
(Vent on
Stop)
4
(Vent on
Stop)
Off
Off
Off
No optional vent selected: Vent when EXT pump speed
falls to below 50% of full speed after Stop is selected.
Off
Off
On
Vent on Stop.
Off
On
Off
Vent on Fail and Vent when EXT pump speed falls to
below 50% of full speed after Stop is selected.
Off
On
On
Vent on Fail and Vent on Stop.
On
*
*
Factory setting.
Off
Off
Vent on Axial Emergency
†
and Vent when EXT pump
speed falls to below 50% of full speed after Stop is
selected.
†
Only available on the EXC300M Controller.
On
Off
On
and Vent on Stop.
On
On
Off
and Vent on Fail and
Vent when EXT pump speed falls to below 50% of full
speed after Stop is selected.
On
On
On
and Vent on Fail and
Vent on Stop.