Controls and indicators (EXC120/300/300M only)
Note:
In addition to the controls and indicators on the front panel, the EXC120, EXC300 and EXC300M Controllers
have an electrical supply isolator on the rear panel (see Figures
The controls and indicators on the front panel of the EXC120, EXC300 and EXC300M Controllers are shown in Figures
and
Start/Stop
Use this latching button to alternately Start or Stop the EXT pump. A LED on the button goes on when EXT Start has
been selected. Note that:
z
The button is wired in series with the Start/Stop input on the logic interface (see
). The EXT
pump will not start until Start is selected from this front panel button and the Start/Stop input is closed. The
Controller is supplied with a wire link to close the Start/Stop input.
z
If the LED on the button is on, it only means that Start has been selected; it does not necessarily mean that
the pump has successfully started.
z
The Controller delays the Start operation until the Interlocks are closed (refer to
); therefore,
the EXT pump will not necessarily start to operate when you select Start. If you select Start and the
Interlocks are not closed, the LED on the button will flash.
This button is also used to reset the Controller after Fail or Axial Emergency condition (see
).
Standby
Use this latching button to select pump Standby mode at any time. In Standby, the rotational speed of the EXT pump
is reduced to 70% of its full rotational speed. Selection of Standby prolongs the life of the pump bearings. Note that:
z
The button is wired in parallel with the Standby input on the logic interface (see
) and you can
use either the button or the input to select Standby.
z
The LED on the button is on when Standby is selected (either by the button or by the Standby input).
Heater
Use this switch to switch on the electrical supply to the bakeout band. Note that the Controller only switches on the
electrical supply to the bakeout band when the EXT pump speed reaches the Normal speed setpoint (see
The LED on the switch is on when the electrical supply to the bakeout band is switched on; this LED can therefore be