Maintenance
After each inspection, it is recommended that:
•
the R-Scan Array scanning head is dried after every use.
•
any dirt or debris are removed from the wheels.
•
the system is packed into the transit case with the umbilical assembly.
Handle replacement
The handle contains the sealed electronics and encoder. To replace:
1.
Remove the waterbox from the scanner.
2.
From beneath remove the two pozi screws securing the dust cover to the front handle.
Figure 10: Front handle dust cover screw location
3.
Remove the four socket cap screws from the front of the handle, two either side of the
wheels.
Figure 11: Front handle front bolt location
4.
Remove the two countersunk screws from the rear of the handle.
Figure 12: Front handle rear bolt location
5.
Lift the handle to separate it from the chassis.
6.
Install the new handle and secure with all screws.