24
Technical Support
For technical support, please contact:
Troubleshooting
Problem
Possible Cause
Action
Encoder not
functioning
Instrument setup
Refer to the instrument’s documentation
regarding encoder setup / calibration.
Encoder cable
Check the encoder cable plug is making
full contact with the corresponding socket.
Check encoder cable plugs are free from
dirt / debris.
Encoder
Encoder is integrated into the handle. Fit
replacement the handle.
Handle
Check the handle is securely fastened to
the chassis.
Buttons not
performing
function
Connection issue
Button is integrated into the handle. Fit
replacement the handle.
Insufficient
probe contact
Both probe lock levers are
not engaged
Ensure both probe lock levers are engaged
to clamp the water box
Water box height
adjustment
Release both probe lock levers on the
scanner and adjust height of water box so
that foam seal is in contact with external
surface. The foam seal should be slightly
compressed.
Plunger wear on the probe lock lever.
Replace the plunger
Wheel Brake not
working
Worn brake shoes
Replace the brake shoes
Incorrect wheels fitted
Using brake when low friction wheels fitted
to scanner. Fit high friction wheels.