Chapter 4 — The QuadStat
43
The Online Indicators
Along the lower edge of the QuadStat front panel are a series Online
indicators,
. When lit, they indicate that the software (such
as EChem, Chart or Scope) has located and initialised that QuadStat
channel. If the light does not go on when the software is run, check that
the QuadStat is properly connected. If there is still a problem, please
refer to Appendix B
The Overload Indicators
Also along the lower edge of the front panel are the Overload
indicators,
. When lit, these indicates that the QuadStat has
overloaded, which usually occurs because it has gone out of
compliance because of an open circuit (such as an unconnected or
faulty electrode), or the resistance is too high in the electrochemical cell.
High resistances can be often be encountered when electrodes are
fouled by the products of electrolysis reactions. The QuadStat tries to
compensate by increasing the compliance potential (that is, the
potential between the auxiliary and working electrodes). If the
compliance voltage exceeds specification, about 11 V, potential control
of the cell is lost and drifting, or oscillation, of the signal can be seen.
Any data collected during this period is unreliable and should be
discarded.
The QuadStat Overload indicators will remain lit once an overload has
occurred — they will be reset once the scan has finished.
If an overload indicator comes on repeatedly, and your connections are
good, then try bringing your electrodes closer together, and/or
increasing electrolyte concentration, and/or modifying your
Color
Electrode
Yellow
Reference
Green
Working
Red
Auxiliary
Table 4–1
Color-coding on the
leads of the electrode
cables.