eDAQ Potentiostats
If an overload indicator comes on repeatedly, and your connections are
good, then try bringing your electrodes closer together, and/or
increasing electrolyte concentration, and/or modifying your
experimental conditions to avoid fouling of the electrodes. Redesigning
your electrochemical cell may be necessary. Normally electrochemical
cells are designed to keep the reference and working electrodes very
close together, however, when a potential overload occurs, you also
4 mm socket,
ground connection
DB-9 pin, I
2
C connectors
20-pin socket for screw
terminal adaptor,
Input and output signals
20-pin terminal adaptor.
Push firmly into socket
Figure 4–2
The QuadStat back
panel.
Figure 4–3
The 20-pin screw
terminal adaptor.