48
eDAQ Potentiostats
software, to connect the QuadStat to other
e-corder
input channels in
which case the description that follows would change accordingly).
When using EChem software,
e-corder
Input 1 is always set to be the
current signal (the I channel), and
e-corder
Input 2 is automatically set
to be the potential signal (the E channel). Thus when using EChem
software connect the QuadStat Channel 1 current signal (I Out) to
e-corder
Input1, and the QuadStat Channel 1 potential signal (E Out) to
e-corder
Input 2. Other QuadStat channels remain unconnected.
To record the applied potential signals (E Out) of a QuadStat with
Chart and Scope software, first configure the settings of the standard
Channel 2
Channel 4
Channel 3
Channel 1
164
QuadStat
RE
RE
RE
AE
AE
AE
AE
Channel 2
Channel 4
Channel 3
Channel 1
164
QuadStat
RE
RE
RE
AE
AE
AE
AE
Channel 1 auxiliary,
reference and working leads
to be connected to AE
1
, RE
1
,
and WE
1
Channel 2 auxiliary and
reference leads connected
together
Channel 2
working lead to
be connected to
second WE
Channel 1 auxiliary,
reference and working leads
to be connected to AE
1
, RE
1
,
and WE
1
Channel 2 reference and working leads to
be connected to RE
2
, and WE
2
for
accurate monitoring of potential at WE
2
.
Note that overload light will be on.
Figure 4–8
These alternative
configurations are
equivalent to a
bipotentiostat.