Acclarix LX4 Diagnostic Ultrasound System User Manual Utilities
- 124 -
Caution
1. For security consideration, a password with high secure intensity is suggested for
each user account, and the password should be changed periodically.
2. Please keep your user name and password safe.
3. Change the passwords of two pre-defined administrator users:Admin1 and Admin2
immediately when you get the ultrasound system.
4. Always enable the function of user login with password protection.
9.2 Connectivity
The Connectivity screens support configuration of network access and services. It is divided into
3 tabs:
TCP/IP
: Configures access to the network from the system.
DICOM
: Configures network DICOM services.
Network Store
: Configures non-DICOM storage to a network via FTP.
WARNING
1. Please build a secure network environment for your ultrasound system to protect the
patient data being attacked during network transmissions.
2. Patient data transmitted by the system is not encrypted. Please ensure the physical
security of the network by trying the following measures:
a) Install IDS(Intrusion Defense System) and IPS(Intrusion Protection System)
software in the network environment where your ultrasound systems are used.
b) Use a network that is isolated from the external environment by a firewall.
c) Use the WPA2 protocol for authentication and encryption of wireless network,
and a password with high secure intensity is required.
d) Do not connect the ultrasound system and non-medical devices to a same
network.
e) Set a login password with high secure intensity at your FTP server side. Keep the
login user name and password of FTP server safe. The leak of user name and
password may lead to patient information and data being disclosed.
3. Anti-virus software is not installed on this device. Please take anti-virus measures in
the network environment where your ultrasound systems are used.
9.2.1. TCP/IP
This screen configures access to a network. Access can be either wired or wireless.