11 EMC Information - Guidance and Manufacture
’
s
Declaration
11.1 Electromagnetic Emissions - for all EQUIPMENT and
SYSTEMS
Guidance and manufacture
’
s declaration
–
electromagnetic emission
The
VE-300 Veterinary Electrocardiograph
is intended for use in the electromagnetic environment specified below.
The user of the
VE-300 Veterinary Electrocardiograph
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 1
The
VE-300 Veterinary Electrocardiograph
uses RF
energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The
VE-300 Veterinary Electrocardiograph
is suitable
for use in all establishments other than domestic
and those directly connected to the public
low-voltage power supply network that supplies
building used for domestic purposes.
11.2 Electromagnetic Immunity - for all EQUIPMENT and
SYSTEMS
Guidance and manufacture
’
s declaration
–
electromagnetic immunity
The
VE-300 Veterinary Electrocardiograph
is intended for use in the electromagnetic environment specified below. The
user of
VE-300 Veterinary Electrocardiograph
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
4 kV contact
±
8 kV air
It is recommended the use of
antistatic materials. If floor are
covered with synthetic material,