VE-300 Veterinary Electrocardiograph User Manual
-35-
RECORD SPEED
In the MANUAL/RHYTHM mode,
RECORD SPEED
can be set as 5, 6.25, 10, 12.5, 25 or
50mm/s.
In the AUTO/OFF/USBPRT mode,
RECORD SPEED
can be set as 25 or 50mm/s.
RECORD LENGTH
Short
form means that each lead group will be recorded about 2.5 seconds.
Medium
form means that each lead group will be recorded about 5 seconds.
Long
form means that each lead group will be recorded about 7.5 seconds.
Longest
form means that each lead group will be recorded about 10 seconds.
Note
: When
RECORD FORMAT
is
1Ch
, only
Short
form is supported in
RECORD
LENGTH
.
5.8.3 Record Output Settings
AVERAGE TEMPLT
When AVERAGE TEMPLT is
2Ch+1R
, limb leads will be recorded in 3 groups of 2, with
one rhythm lead at the bottom of recoding paper.
When AVERAGE TEMPLT is
3Ch
, limb leads will be recorded in 2 groups of 3.
When AVERAGE TEMPLT is
Off
, there will be no average template when recording.
RR ANALYSIS
When
RR ANALYSIS
is
On
, RR Analysis results, including RR Interval measurement
information, RR Histogram and RR Trend Chart, will be recorded after rhythm wave is
recorded in the RHYTHM mode.
When
RR ANALYSIS
is
Off
, there will be no RR Analysis results after rhythm wave is
recorded in the RHYTHM mode.
Average Template :
2Ch+1R
Measurement : Off
Interpretation :
Off
RR Analysis :
On