VE-300 Veterinary Electrocardiograph User Manual
-34-
to turn on the function and select
Off
to turn off.
EMG FILTER
EMG FILTER suppresses disturbances caused by strong muscle tremor. The cutoff frequency
is user-defined at
25Hz, 35Hz
or
45Hz
. Select
Off
to turn off the function.
DFT FILTER
DFT FILTER greatly reduces the baseline fluctuations without affecting the ECG signal. The
purpose of this filter is to keep the ECG signals on the baseline of the printout. The setting
value is the low limit of the frequency range, including
0.05Hz
,
0.15Hz
,
0.25H
z,
0.5Hz
, and is
normally set to
0.15
Hz.
LOWPASS FILTER
LOWPASS FILTER
restricts the bandwidth of input signal. The cutoff frequency is user
defined at
150Hz
,
100Hz
or
75Hz
. All the input signals whose frequency is higher than the
setting cutoff frequency will be attenuated.
5.8.2 Record Settings
RECORD FORMAT
When
RECORD FORMAT
is
3Ch
, limb leads will be recorded in 2 groups of 3.
When
RECORD FORMAT
is
3Ch+1R
, limb leads will be recorded in 2 groups of 3, with
one rhythm lead at the bottom of recording paper.
When
RECORD FORMAT
is
1Ch
, all leads will be recorded one by one in a sequence.
When
RECORD FORMAT
is
1Ch+1R
, all leads will be recorded one by one in a sequence,
with one rhythm lead at the bottom of recoding paper.
RECORD GRID
When
RECORD GRID
is
On
, the dashed grids which are 5
mm by 5
mm will be recorded on
the paper.
When
RECORD GRID
is
Off
, dashed grids will not be recorded on the paper.
Record Format :
3Ch
Record Grid : Off
Record Speed :
25mm/s
Record Length :
Short