VE-300 Veterinary Electrocardiograph User Manual
-49-
7 Technical Specifications
Safety
Standards
IEC 60601-1:1988+A1+A2, EN 60601-1:1990+A1+A2,
IEC/EN 60601-1-2:2001+A1, IEC/EN 60601-2-25, ANSI/AAMI EC-11
Anti-electric-shock type:
Class
І
with internal power supply
Anti-electric-shock degree:
Type CF
Degree of protection against harmful
ingress of water:
Ordinary
equipment
(Sealed
equipment without liquid proof)
Disinfection/sterilization method:
Refer to the user manual for details
Degree of safety of application in the
presence of flammable gas:
Equipment not suitable for use in
the presence of flammable gas
Working mode:
Continuous operation
Classification
EMC:
Group
І
, type A
Dimensions
288mm
×
210mm
×
70mm
Weight
About 2.5kg
Display
192
×
64 dot single color LCD Screen
Transport/Storage
Working
Temperature:
-20
℃
~55
℃
5
℃
~40
℃
Relative
Humidity:
25%~93%
non-condensing
25%~80%
non-condensing
Environment
Atmospheric
Pressure:
700hPa ~1060hPa
860hPa ~1060hPa
Rated input voltage =100V~115V/220V~240V
Rated frequency = 50/60Hz
Mains Supply:
Rated input power = 35VA
Rated voltage = 14.8V
Rated capacity =
2000mAh /2200mAh /2400mAh
Charge mode: Constant current/voltage
Power Supply
Built-in Lithium Battery
Pack:
Charge current (standard) = 0.2C
5
A (320mA)