VE-300 Veterinary Electrocardiograph User Manual
-16-
Definition of corresponding pins:
Pin
Signal
Pin
Signal
1
GND
4
GND
2
GND
5
ECG Signal (input)
3
GND
6
ECG Signal (output)
4) USB Interface
Definition of corresponding pins:
Pin
Signal
Pin
Signal
1
VBUS
3
D+
2
D-
4
GND
WARNING
: Only the USB equipments recommended by EDAN can be
connected to the USB interface.
WARNING
:
♦
Accessory equipment connected to the analog and digital interfaces must
be certified according to the respective IEC/EN standards (e.g.
IEC/EN60950 for data processing equipment and IEC/EN 60601-1 for
medical equipment). Furthermore all configurations shall comply with the
valid version of the standard IEC/EN 60601-1-1. Therefore anybody, who
connects additional equipment to the signal input or output connector to
configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC/EN 60601-1-1.
If in doubt, consult our technical service department or your local distributor.
♦
The summation of leakage current should never exceed leakage current
limits while several other units are used at the same time.