10
with a gentle movement.
To fold down the armrest grasp the upholstery and move it sideways towards the feet end and the
armrest will automatically unlock, then push it down.
This table model is equipped with ARM SUPPORTS at the headrest allowing to place the arms alongside
head when in prone position.
To adjust the arm supports height, simply loosen the wing nut and slide the arm support to the required
position. Tighten the wing nut back on to lock the arm support.
The arm supports are also depth-adjustable and removable. Simply loosen the wing nut beneath the
headrest, grasp the arm support and pull outwards. Then tighten the wing nut to lock the arm support.
*Caution! The patient must not lean on the armrest as it could split following an excessive strain.
3) MAINTENANCE
For safety reasons, DISCONNECT THE TABLE from the electrical supply before the table
cleaning, repair or servicing.
a) Metal FRAME
For a proper operation of the table, you should do the dusting regularly with a cloth.
We also recommend to periodically grease the axles at the moving parts with a liquid grease spray. Lay
the couch on its side and take off the white caps protecting the nuts at the axles of the framework. Slightly
unscrew the nuts and spray the product around the loosen nuts. Leave to settle in this position for 12
hours. Tighten the nuts and put the protective white caps back on. Lay the table on the other side and
repeat.
Periodically check the couch visually for any signs of damage and make sure that all bolts, nuts and other
fasteners are tight.
* Please note that the repeated exposure of the table to high levels of humidity is not recommended.
b) Electric MOTOR
Keep this piece free from any dust or grease.
Absolutely avoid spilling liquids on the motor because they could cause severe damage to the electric
components despite the motor protection.
THE HANDLING OF THE ELECTRIC MOTOR IS TOTALLY FORBIDDEN. Any manipulation will
invalidate the guarantee.
A non-stop blinking operation of the foot rail or any other height adjustment mechanism could activate the
thermal protection of the motor due to an overheating.
For further information, go to annex “FREQUENTLY ASKED QUESTIONS ABOUT MOTORS”.
Examine the table periodically for any signs of damage or malfunction and contact immediately the reseller
or Ecopostural for service/ spare parts.
c) Section adjustment MECHANISMS
Keep the mechanisms free from any dust or grease.
THE HANDLING OF THE GAS SPRING AND/OR THE ELECTRIC MOTOR IS TOTALLY FORBIDDEN.
Any manipulation will invalidate the guarantee.
If any malfunction or failure is suspected, contact immediately the reseller or Ecopostural.
d) UPHOLSTERY
A wet cloth should be enough to do the dusting. Anyway, if the leather is not properly clean you can use a
smooth sponge soaked with neutral soap water. Dry immediately afterwards. To prevent moisture from
entering the seams, do not wash with abundant water.
It is highly recommended to use a towel cover like the A4469, A4459 or A4460 to protect the upholstery
from sweat, massage creams or oils.
Use soapy water to clean the stains. Avoid using alcohol, any solvent or stain remover as the vinyl
could be seriously damaged.
The use of infrared lamps might cause a cracking of the artificial leather. You should then put a cotton