43
Figure 6
4. Measuring Current
To measure current use the input terminals in the front of the calibrator. Select the mA
function on one of the displays. Current is measured in mA and percentage of range.
The range on the calibrator is set to 0% at 4 mA and 00% at 20 mA.
For example:
If the current measured is displayed as 75% then the mA value is 6 mA.
Figure 7
Model BPPA
Pressure Module
Adapter
Pressure Module
Device
under
Test
4 to 20mA
P
P2
30V
24mA
max
Содержание CP 400
Страница 1: ...CP 400 420 Bedienungsanleitung Operating Instructions Notice d utilisation ...
Страница 33: ...64 15 Menu Map ...