Page 15
ECOM-P
(Adjustment
<F1>
for Yes /
<F4>
for No):
- Adjusted of selected printer:
„YES“
= ecom-P
„NO“
= HP 82240B
Eta(C)
(
<F1>
for YES /
<F4>
for NO):
- Efficiency calculation with or without condensation gain
6. Control
The electrochemical sensors used for the gas analysis are submitted
to a wearing process and consequently do age. Along the operation
period, they alter their output values depending on the gas concen-
tration, flow duration and soiling degree of the measured gases. The
programme controls the sensors and corrects drifts. If drifts and the
correlated measurement errors increase, an error message is dis-
played. In this case the corresponding sensor must be changed by
an authorised service centre.
The control menu informs about the current status values for the
sensors and on 2 further pages (scroll display page via cursor key)
about:
- the battery voltage (charging state)
- the phone number of the next service centre
- the serial number
- the software version
- the operation hours
- the date of the last service
- the number of CO switch offs
- the number of errors occurred
O2
1034 mV
CO
-10 mV
Batt
6.42 Volt
Further info with
↑↓
!
Tel.No. 02371/945-5
Ins.No. CN-0009 V1.1
Oper. hours
7.39
Service date 23.06.04
CO switch offs
0
Error counter 00
Further info with
↑↓
!
Tel.No. 02371/945-5
Ins.No. CN-0009 V1.1