Change fine dust filter (condensate trap basis version):
Release the upper cap. Release the fine dust filter and check its
state. It should be replaced once the soiling degree in the gas flow
area shows a grey colour (ca. number 3 on soot comparison scale).
The safety filter should also be changed when the soiling degree
corresponds to number 3 of said comparison scale.
Empty condensate trap (condensate trap basis version):
The condensate collects the humidity of the sampled gas. Check
regularly how much liquid has been gathered. Empty the trap in time
and avoid humidity to enter the instrument. Release the lower cap to
empty the trap. Pour out the condensate. Reassemble the trap.
Make sure that condensate never
enters the instrument (otherwise sensors
will get defective)!
Condensate trap without monitoring
(Basis version)
Condensate trap
Safety filter
Fine dust filter