87
Table 17: RTD Read and Source
RTD Type
Range (°C)
Accuracy (°C)
Ni120 (672)
-80.0 ... 260.0
0.2
Cu10
-100.0 ... 260.0
1.4
Cu50
-180.0 ... 200.0
0.4
Cu100
-180.0 ... 200.0
0.3
YSI400
15.00 ... 50.00
0.1
Pt100 (385)
-200.0 ... 100.0
0.2
100.0 ... 300.0
0.3
300.0 ... 600.0
0.4
600.0 ... 800.0
0.5
Pt200 (385)
-200.0 ... 100.0
0.8
100.0 ... 300.0
0.9
300.0 ... 630.0
1.0
Pt500 (385)
-200.0 ... 100.0
0.4
100.0 ... 300.0
0.5
300.0 ... 630.0
0.6
Pt1000 (385)
-200.0 ... 100.0
0.2
100.0 ... 300.0
0.3
300.0 ... 630.0
0.4
Pt385-10
-200.0 ... 100.0
1.4
100.0 ... 300.0
1.6
300.0 ... 600.0
1.8
600.0 ... 800.0
2.0
P1385-50
-200.0 ... 100.0
0.4
100.0 ... 300.0
0.5
300.0 ... 600.0
0.6
600.0 ... 800.0
0.7
Pt100 (3926)
-200.0 ... 100.0
0.2
100.0 ... 300.0
0.3
300.0 ... 630.0
0.4
R100 (391 6)
-200.0 ... 100.0
0.2
100.0 ... 300.0
0.3
300.0 ... 630.0
0.4
Read Accuracy is based on 4-wire input. For 3-wire input add ± 0.05Ω assuming all three