86
Table 15: Thermocouple Measurement / Source
Range
Accuracy(% of reading ± digit)
Read (mV)
-10.000mV - 75.000mV
0.02% ± 10
Source (mV)
-10.000mV - 75.000mV
0.02% ± 10
Maximum current output in voltage ranges is 1mA with an output impedance of 1Ω
Table 16: Thermocouple Read and Source
TC Type
Range (°C)
Accuracy (°C)
J
-210.0 ... 0.0
0.4
0.0 ... 800.0
0.2
800.0 ... 1200.0
0.3
K
-200.0 ... 0.0
0.6
0.0 ... 1000.0
0.3
1000.0 ... 1372.0
0.5
T
-250.0 ... 0.0
0.6
0.0 ... 400.0
0.2
E
-250.0 ... -100.0
0.6
-100.0 ... 1000.0
0.2
R
0.0 ... 1767.0
1.2
S
0.0 ... 1767.0
1.2
B
600.0 ... 800.0
1.2
800.0 ... 1000.0
1.3
1000.0 ... 1820.0
1.5
C
0.0 ... 1000.0
0.6
1000.0 ... 2316.0
2.3
XK
-200.0 ... 800.0
0.2
BP
0.0 ... 800.0
0.9
800.0 ... 2500.0
2.3
L
-200.0 ... 0.0
0.25
0.0 ... 900.0
0.2
U
-200.0 ... 0.0
0.5
0.0 ... 600.0
0.25
N
-200.0 ... 0.0
0.8
0.0 ... 1300.0
0.4
All TC errors include CJC errors
CJC errors outside of 23 ± 5 °C is 0.05°C/°C (In °C add .2 for cold junction compensation error.)
<-