6.1 CLEANING
Before cleaning the PROCAM HD, disconnect the power cable from the device and from the
electric power supply.
6.1.1 EXTERNAL SURFACE
To clean the outer surfaces of the device only use cotton or cloth dampened in mild detergent or
soap. Only use adequate cleaning products prepared for electrical equipment following the products
instruction.
While cleaning, be careful not to let water or humidity get into the equipment.
Never use flammable products, especially alcohol.
Do not turn the equipment on until one (1) hour after cleaning process.
We recommend cleaning the external surface of the equipment every 15 days.
Do not sterilize any parts of the device.
Do no immerse the device or its accessories.
The cleaning, disinfection and sterilization of accessories not provided by ECLERIS must be
performed following the instruction provided by their manufacturer.
6.2 ELECTRICAL CONNECTIONS
Permanently check the condition of all the electrical components.
Check that there are no:
•
Cables in bad conditions; cut or torn.
•
Bad connection of the cable with the socket.
•
Rusty pins in the connectors.
•
Disconnected equipotential ground connection (in places with equipotential ground).
6.3 CHANGING THE FUSE
To avoid a waste of time for unavailability, we recommend having spare fuses on hand.
1. Disconnect the power cable from the electric current and from the device.
2. Next to the power cord connector there are two plastic knobs where the fuses are housed.
Manually turning these knobs in the direction indicated by the arrows, both handles can be
removed, one at a time, to replace damaged fuses.
3. Remove the damaged fuses and replace them with the spare fuses, to do this use the same
value of the replaced fuses.
Only use fuses as shown at the rear side of the PROCAM HD console and certified by
IEC60127.
4. Screw again the knobs.
5. Connect the power cord.