If any abnormality is detected before turning on the equipment, please contact the technical
service of your local distributor or your nearest ECLERIS office before using it.
5. OPERATION
In this section there are detailed procedures to turning on and verifying the equipment functions.
We recommended following them after the installation.
Only medical, paramedical or health personnel should work this equipment.
Do not sterilize the device. For use in the operating room use a sterile sheath.
1. Press the power button in front of the PROCAM HD console. The LEDs of the PROCAM HD
camera will blink for few seconds and then they will turn off. It indicates that the PROCAM HD
camera head is being powered.
2. Verify that the video monitor is ON and in Video mode through its HDMI/DVI / VGA or CVBS
input.
3. Proceed with the white balance.
4. Adjust the picture on the monitor according to your preferences of Brightness, Contrast,
Saturation, Color, HUE, Gamma.
5. Adjust the functions according to your preferences: SPECIALTY SELECTION.
6. Once the procedure is over, turn the device off in order to maximize the system lifetime and
energy saving.
5.1 CONSOLE FUNCTIONS
5.1.1 WHITE BALANCE
To obtain the white balance, focus on a completely bright and white image lit with the maximum
intensity and press the White Balance button for approximately 3 seconds. Release the
button and continue focusing on the white image till the icon of the padlock on the screen
get closed (approximately 2 seconds). This set up will be stored in the memory even if the
camera is off.
While the White Balance function is being performed, the icon of the padlock on the screen will
indicate that the function is activated.
If the white balance is well adjusted, this guarantees the reproduction of natural colors.
Perform a new white balance every time the light source is changed.
Perform the White Balance with the same light source that you will use in the intervention.
5.1.2 SPECIALTY SELECTION FUNCTION
The Specialty Selection function allows you to select different predefined image
configurations, associated to different applications and endoscope types.