Ecleris CAVIX Скачать руководство пользователя страница 15

USER MANUAL

– CAVIX 

 

Page 

15

 of 

16

 

INMANUUIN185 Rev.2019.04.01 / ENG 

35 - 75 % relative humidity  
700 - 1060 HPa 

Storage and transport conditions 

30 -  45 ºC 
10 - 95 % relative humidity 
500 - 1060 HPa 

 

10. 

TROUBLESHOOTING 

The following table indicates the most common problems and their solutions. Please consult this 
table if you have any inconvenience. 

PROBLEM 

PROBABLE CAUSE 

SOLUTION 

The  equipment  will 
not turn on 

It is not connected to the 
electric power supply. 

Connect  the  cable  to  the  electric 
power supply. 

The  power  cable  is 
damaged or cut. 

Replace the power cable. 

The  plug  is  in  bad 
conditions. 

Replace the plug. 

The  fuse  is  damaged  or 
burnt. 

Replace the fuse. 

The  green  pilot  light 
on  the  switch  turns 
on  but  the  screen 
does not turn on. 

Touch  screen  or  board 
is damaged. 

Contact  your  nearest  Technical 
Service. 

Emission 

is 

not 

functioning. 

The  output  selected  is 
incorrect  or  the  intensity 
is too low. 

Select 

the 

correct 

output 

pressing  the  button  (8)  on  the 
front panel. 
Check the programmed intensity. 

Defective applicator. 

Check 

the 

estate 

of 

the 

applicator 

cable 

and 

its 

connection  to  the  equipment. 
Replace the applicator. 

 

If the inconvenience can’t be solved, please contact your local distributor or the nearest ECLERIS 
office and ask for technical service.  

 

11. 

TECHNICAL SERVICE 

CAVIX  systems  should  be  repaired  and/or  maintained  exclusively  in  ECLERIS  authorized 
installations. 

ECLERIS,  will  provide  if  required,  circuits,  drawings,  list  of  components,  descriptions  and 
instructions  for  calibration  or  other  useful  information  required  by  qualified  trained  technical 
personnel, to repair the parts of the device marked as reparable by the manufacture. 

We  recommend  using  the  following  procedure  to  expedite  repair  service  and  reposition  of 
products under guarantee. 

Contact the technical service department of your local distributor or nearest ECLERIS office.  

Содержание CAVIX

Страница 1: ...can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be modified or broaden without previous notification without it r...

Страница 2: ...ALLATION AND SET UP 9 4 1 UNPACKING 9 4 2 POSITIONING 9 4 3 REVISION 9 5 OPERATION 9 6 TREATMENTS 11 6 1 BEFORE STARTING THE TREATMENT 11 6 2 GENERAL PRECAUTIONS 11 6 3 CONTRAINDICATIONS 11 6 4 COMPLI...

Страница 3: ...RIS will not assume any liability for damages that may occur if the device is used without following the instructions and or ignoring the following advice Keep the device out of children reach Do not...

Страница 4: ...permanent review If the system is sold or donated the user manual must be deliver with the device If by any of the following Inadequate use of the equipment Failure to use the instruction manual Bad m...

Страница 5: ...ng liquid on the equipment The manufacture will not be held responsible for any use of the equipment different from that which it is intended The use of commands controls or procedure sequences that a...

Страница 6: ...e to a surgical treatment of the localized fatty layers Furthermore thanks to the synchronization system CAVIX ensures a calibrated frequency of ultrasound emission guaranteeing a safe treatment respe...

Страница 7: ...t upon the fatty tissue 3 3 INTENDED USER The equipment is intended to be used by dermatologists or beauticians with specific training in the technique of cavitation 3 4 COMPONENTS The CAVIX includes...

Страница 8: ...INMANUUIN185 Rev 2019 04 01 ENG 3 Applicator connector 4 Pedal connector 5 Power cable conenector with fuse holder 6 Power switch 7 Air vent CAVITATION APPLICATOR CAVITATION APPLICATOR SUPPORT FOOTSW...

Страница 9: ...n or if it is on turn it off immediately and consult one of our authorized assistance centers 4 3 REVISION Once the installation is complete and before turning on the CAVIX we recommend carrying out t...

Страница 10: ...the RANGE value respectively the indicator 2 will show the programmed value This value indicates the range of frequencies emitted only when the SWEEP mode is ON The range of emitted ultrasound freque...

Страница 11: ...xclusively by operators properly trained in the use of cavitation equipment and only after reading this manual completely The information provided in this section does not aim at covering all the prob...

Страница 12: ...ld communicate the sensations experimented and the variations of these sensations To be explicit this is not a treatment with definite results because that would depend on eating habits and possible c...

Страница 13: ...ns Periodic cleaning of all external components Clean applicators Check connections Annual technical verification of the equipment For any type of maintenance work or revision contact your local distr...

Страница 14: ...aving a spare fuse on hand 1 Disconnect the power cable from the electric current and from the device 2 Under the connector for the power cable there is a small rectangular plastic piece where the fus...

Страница 15: ...ing The output selected is incorrect or the intensity is too low Select the correct output pressing the button 8 on the front panel Check the programmed intensity Defective applicator Check the estate...

Страница 16: ...CAVIX has passed our Quality Control and no defects have been found in its manufacture therefore guarantees proper working order under adequate conditions that are indicated in this manual The guaran...

Отзывы: