User manual SEFELEC 5x Series
179
v1.06
*CLS
Positions all standardized registers in the state in which the user finds them at power
on.
*RST
Performs a reset of the device.
WARNING: The unit returns to Local mode, regain control of the device with the
REM command before any other command.
18.1.3.
STANDARDIZED QUERIES
*STB?
Returns to the NR1 format and in the hexadecimal base the value of the "STB" register
(e.g.: "#H80").
b0
0 = safety loop open
1 = safety loop closed
b1
0 = no error 1 = error (incorrect voltage, etc.)
b2
0 = End of test 1 = Test In Progress
b3
0 = Test FAIL 1 = Test PASS
b4
Not used
b5
Or Logic of the bits of the ESR Register
b6
Or Logic of the bits of the STB Register
b7
Not used
*SRE?
Returns to the NR1 format and in the hexadecimal base the value of the "SRE" register
(STB register mask)
*ESR?
Returns to the NR1 format and in the hexadecimal base the value of the register "ESR"
b0 to b3 Not used
b4
1 = type 2 dialogue error (digital value outside limits, command out of context)
b5
1 = type 1 dialogue error (incorrect syntax)
b6
Not used
b7
1 = Switching on the power
Note: The bits are reset after reading by the *ESR? command.
Содержание SEFELEC 5 Series
Страница 47: ...User manual SEFELEC 5x Series 46 v1 06 Proceed in the same way for setting the time and validate by OK...
Страница 131: ...User manual SEFELEC 5x Series 130 v1 06...
Страница 159: ...User manual SEFELEC 5x Series 158 v1 06 Diagram No 2 Launch of an insulation test Memory 5 Test result NG...
Страница 160: ...User manual SEFELEC 5x Series 159 v1 06 Diagram No 4 Wiring of the PLC option with an external power supply...
Страница 205: ...User manual SEFELEC 5x Series 204 v1 06 EU DECLARATION OF CONFORMITY...
Страница 206: ...User manual SEFELEC 5x Series 205 v1 06...
Страница 207: ...User manual SEFELEC 5x Series 206 v1 06...
Страница 208: ...User manual SEFELEC 5x Series 207 v1 06...