76
v1.06
If the measurement is < Rmin or >Rmax, it will be regarded as FAIL
:
The coloured function key indicates the setting that caused the FAIL sanction of the measure-
ment.
Before disconnecting the measured component, press the Stop button to con-
firm the stoppage of the high voltage generation and discharge the component: the de-
vice then performs a check of the residual voltage up to a value of 30 VAC or DC
This action returns you to the initial measurement screen either to change the settings or to per-
form a new measurement by pressing the Start button.
STOP