User manual SEFELEC 5x Series
156
v1.06
CTRLOUT (Output C7-5)
Validation contact for the taking of control of the device by the PLC.
EOT (Output C7-7)
End of Test contact
PLCPASS (Output C7-8)
Test result contact: PASS
PLCFAIL (Output C7-6)
Test result contact: FAIL
ERROR (Output C7-9)
Indicates whether an error occurred during the test; outputs SR0 to SR3 give the type of
error
SR0 to SR3 (Inputs C8-5,6,7,8)
Coding of the type of error according to the following table:
SR3 SR2
SR1
SR0 Type of error
0
0
0
0
Open Safety loop
0
0
0
1
S
aturation
0
0
1
0
Over range
0
0
1
1
Charge
0
1
0
0
Voltage error
-
-
-
-
…
1
1
1
1
….
IN_OUT_0-10V_0 (Output C6-1)
Analogue output 0-10VDC, configurable
IN_OUT_0-10V_1 (Output C6-2)
Analogue output 0-10VDC, configurable
IN_0-10V_2 (Input C8-9)
Analogue input 0-10VDC, configurable
IN_ 0-10V_3 (Input C7-3)
Analogue input 0-10VDC, configurable
Содержание SEFELEC 5 Series
Страница 47: ...User manual SEFELEC 5x Series 46 v1 06 Proceed in the same way for setting the time and validate by OK...
Страница 131: ...User manual SEFELEC 5x Series 130 v1 06...
Страница 159: ...User manual SEFELEC 5x Series 158 v1 06 Diagram No 2 Launch of an insulation test Memory 5 Test result NG...
Страница 160: ...User manual SEFELEC 5x Series 159 v1 06 Diagram No 4 Wiring of the PLC option with an external power supply...
Страница 205: ...User manual SEFELEC 5x Series 204 v1 06 EU DECLARATION OF CONFORMITY...
Страница 206: ...User manual SEFELEC 5x Series 205 v1 06...
Страница 207: ...User manual SEFELEC 5x Series 206 v1 06...
Страница 208: ...User manual SEFELEC 5x Series 207 v1 06...