FIG . 5- To adjust the height of the glide handles, loosen ratchet clamps. Adjust the handles to the 

user’s desired position and tighten the clamps securely.

The resistance cylinders are individually adjustable to provide twelve different resistance settings. Turn 

the cap of the cylinders to the desired resistance. The resistance cylinders can also be disconnected 

by disengaging the pull pin. 

FIG . 6- Adjust the foot plates by loosening the two knobs on each foot plate (for easier access to the 

inside knob, swing the leg assembly forward), so the user’s knees are centered in the knee pads and 

the femur is resting on the seat. Tighten securely.

Caution: Never reposition the foot plates while a user is in the standing position.

 

Caution: Avoid placing fingers in the adjustment slot.

 

 

FIG . 7- Often, proper fit can be achieved by adjusting the foot plate height. When necessary, the knee 

pads are independently adjustable with five adjustments. To adjust the knee pads, remove the bolts 

from the knee pads and adjust to the desired position. Replace the bolts and tighten securely.

FIG . 8- Before transferring, flip up the knee pads to allow easier access. 

Caution: Never make adjustments to the EasyStand while a user is in the standing position.  

Caution: While using the Glider, advise bystanders to keep hands and feet clear of the Glider 

mechanism.

Caution: While using the Glider, keep hands clear of the resistance cylinders’ upper and lower 

mounting points.

FIG. 5

FIG. 7

Adjustments

FIG. 8

!

!

!

FIG. 6

Page 19 - Perfect Fit Guide

Adjustments - Page 4

Perfect Fit Guide

!
!

step one - measure

Before transferring into the Glider, measure for 
approximate (A) seat depth and seat to (B) foot plate 
settings. 

Caution: These adjustments are only to be made while the 
unit is unoccupied.

step two - seat depth

To adjust seat depth, loosen the knob under the 
seat (C). Use measurement (A) and measure from 
the inside of the knee pad to the back. Depress the 
snap button on the adjustment tube while moving 
the back (D). Adjust the seat depth until the proper 
measurement has been reached. Tighten the knob 
(C) after adjusting.

Caution: The following adjustments are only to be made in 
the seated position

step three - back angle

After achieving the desired seat depth, rotate the 
back angle (E) adjustment knob so the (F) seat/back 
angle indicator on the side of the unit is in the green 
area prior to standing.

Note: Repeat steps 2 and 3 for each color adjustment.

 

step four - foot plates

Use the (B) measurement to adjust the foot plate 
height. Loosen the (G) knobs on the foot plates to 
adjust height. Tighten the knobs after positioning. 
The knees should be slightly lower than the hip (H).

step five - positioning

Position the user’s hips as close as possible to (I) 
Glider’s hip pivot. 

step six - knee pad

Slide into knee pads (J) and secure the calf straps 
behind the knees.

(C) 

(D) 

(B) 

(A) 

(F) 

(E) 

(G)

(H) 

(I)

(J)

Inspect your standing frame periodically to ensure it is in safe operating condition.

!

!

Содержание Glider Medium

Страница 1: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Страница 2: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Страница 3: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Страница 4: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Страница 5: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Страница 6: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Страница 7: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Страница 8: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Страница 9: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Страница 10: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Страница 11: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Страница 12: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Страница 13: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Страница 14: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Страница 15: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Страница 16: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Страница 17: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Страница 18: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Страница 19: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Страница 20: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Страница 21: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Страница 22: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Страница 23: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Страница 24: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Страница 25: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Страница 26: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Страница 27: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Страница 28: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Отзывы: