Safety Precautions 

•   If a user does not meet these specifications and has not been properly fitted by a qualified therapist or physician,   

 

Altimate Medical, Inc., does not recommend using the EasyStand. 

  

 

Glider Medium -

 is designed to accommodate most individuals from 4’0”-5’6” (122-168cm) and up to 200 lbs. (90 kg) 

 

Glider Large -

 is designed to accommodate most individuals from 5’0”-6’2” (152-188cm) and up to 280 lbs. (127 kg) 

 

 

 

 

For more detailed specifications visit our website easystand.com

•   Altimate Medical, Inc., recommends consulting with a therapist or physician prior to starting a standing program. 
•   Check the EasyStand periodically to ensure that all nuts, bolts and adjustable parts are tightened securely.
•  Never make adjustments to the EasyStand while a user is in the standing position. 
•   Never reposition the foot plates while a user is in the standing position. Consult with a therapist or physician for 

 

proper foot placement and angle. 

•  It is necessary to inspect the EasyStand Glider at least weekly to ensure it is in safe operating condition. Pay 

 

particular attention for loose hardware. If the EasyStand is equipped with the optional Pow’r up lift please inspect 

 

the electrical cables for any wear and to ensure they are attached properly. Replacement of worn parts, major 

 

adjustments, or any other important corrections should be handled by an authorized dealer or service center. Only 

 

Altimate Medical, Inc., approved replacement parts should be used to ensure safety and performance.

•  If the EasyStand is being used in the home environment, please inspect the unit prior to each use to ensure there is 

 

no damage or unexpected wear to the EasyStand that may have been caused unintentionally by pets, pests or children.  

•   If the unit was exposed to temperatures of less than 50˚F (10˚C), the unit must warm to room temperature before use.
•  Operating Conditions: Temperature: 41˚F (5˚C) -104˚F (40˚C), Relative humidity: 20% - 90% at 86˚F (30˚C),  

 

 

Atmospheric Pressure: 700 to 1060hPa.

•  Storage/Transport Conditions: Temperature: 14˚F (-10˚C) -122˚F (50˚C), Relative humidity: 20% - 90% at 86˚F (30˚C), 

 

Atmospheric Pressure: 700 to 1060hPa.

•  The EasyStand can be used at a maximum altitude of 3,050 meters (10,000 feet).
•  The expected service life of an EasyStand is considered 5 years under normal use conditions. Note: This may vary 

 

based upon use.

•   Use body weight to lower the seat. Do not use excessive force.
•   The EasyStand is for indoor use on level surfaces only.
•   Violently thrusting in the EasyStand may cause it to tip.  
•   Always put the rear casters and/or wheels in the locked position before transferring into the EasyStand. Use these 

 

locks to keep the EasyStand stationary.

•  The EasyStand Glider meets EN12182:1999 (test standard) for strength, durability and tipping stability.
•  Please contact Altimate Medical, Inc., for relevant clinical data and literature, strength, durability and test results as  

 applicable.
•   The EasyStand Glider Electric Lift meets the requirements of IEC 60601-1 and 60601-2.(Pow’r Up Lift Option Only)
•   Altimate Medical products are specifically designed to be used with Altimate Medical accessories and options  

 

Unless otherwise noted, accessories and options from other manufacturers have not been tested by Altimate  

 

Medical and are not recommended for use with Altimate Medical products.

•  Product Modifications made without express written consent (including, but not limited to, modification through the 

 

use of unauthorized parts or attachments) are not recommended and will void the product warranty.   

 Caution: This equipment is designed to accommodate a range of users; therefore the unit has various adjustment 

 

slots and holes. Avoid placing fingers or other body parts in any of these areas.  

 Caution: While in use this equipment has various moving parts, familiarize yourself with your unit and avoid placing 

 

hands or fingers near any of these areas.  

 Caution: Advise bystanders to keep clear of the unit while in use.

!

!

!

Page 21 - Standing

Safety Precautions - Page 2

Contraindications

Standing/weight bearing may not be an option for the following reasons:

• If the user has severe contractures or range of motion issues

• If the user has severe osteoporosis or brittle bone disease

• If the user has postural orthostatic hypotension

• If the user has dislocated joints

 

Caution: Consumers considering a standing program should consult with a qualified 

physician and medical team to determine what indications or contraindications exist. 

Clinicians must use best clinical judgment to evaluate the user’s needs.

 

Implementing a standing program must be determined on an individual basis by the user’s 

team (Physician, Physical or Occupational Therapist, Rehab Technology Professional/Supplier 

and the consumer). Consider all indications and contraindications for each client individually 

to determine if standing/weight bearing is appropriate for them.

To Stand

Caution: Always consult with a physician or therapist before starting a standing program. 

Caution: Before standing, the seat/back angle indicator must be in green. Failure to set the 

back angle in green prior to standing could cause the unit to lock in a standing position, 

causing damage to the unit. Do not use the seat angle dial indicator as a reference while in 

the standing position. It is only to be used as a guide while adjusting the seat depth 

and back angle in the seated position . 

Position feet, the knee support, and secure any straps or side supports. Push the pump 

handle forward with 15-25 long strokes. The seat will slowly rise to the standing position. 

Remember the seat can be stopped anywhere between sitting and standing. 

Caution: Never make adjustments to the EasyStand while a user is in the standing position. 

Return to the Seated Position

Move and hold the pump handle toward the rear until the desired position is reached. The 

seat will slowly lower until the handle is released. If difficulty lowering the seat occurs, apply 

less or more pressure to the handle. If the unit is in the standing position and unoccupied, 

apply moderate pressure to the seat of the unit while holding the pump handle back. Do not 

use excessive force.

Caution: When lowering the unit from standing to sitting, keep all hands and feet clear of 

the mechanism beneath the seat.

Standing

!

!

!

!

!

Содержание Glider Medium

Страница 1: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Страница 2: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Страница 3: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Страница 4: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Страница 5: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Страница 6: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Страница 7: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Страница 8: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Страница 9: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Страница 10: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Страница 11: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Страница 12: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Страница 13: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Страница 14: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Страница 15: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Страница 16: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Страница 17: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Страница 18: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Страница 19: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Страница 20: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Страница 21: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Страница 22: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Страница 23: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Страница 24: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Страница 25: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Страница 26: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Страница 27: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Страница 28: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Отзывы: