Congratulations on your purchase of an EasyStand standing frame. We at Altimate Medical have 
designed the EasyStand with your comfort and happiness in mind.

EasyStand Description 
The EasyStand line of products is intended for individuals who are able to transfer independently or with 
assistance from a caregiver and/or lifting device. It is a useful tool for rehabilitation, home, school, and 
work use. EasyStand standing frames support the user from sitting to the standing position. Various 
options such as support, choice of manual or Pow’r Up Lift may be added to facilitate individual needs. 

Indications for Use 
The EasyStand Glider line of products is intended to assist individuals who have difficulty rising from a 
seated position to a standing position and is indicated for individuals weighing up to 200 lbs. (Medium), 
280 lbs. (Large).

This manual is an integral part of your standing system and should be read completely and 
thoroughly by anyone that will be operating the unit .
 This will ensure the EasyStand is assembled 
properly and used in the most beneficial manner. 

Altimate Medical recommends that an attendant be present when the EasyStand is being used. Should 
any problem arise that you cannot solve by reading this manual, please contact your local supplier or 
call Atimate Medical, Inc., at 800-342-8968 or 507-697-6393.

It is necessary that you inspect your EasyStand Glider at least weekly to ensure it is in safe 
operating condition. EasyStand adjustments that are listed in the product owner’s manual can be 
made by the operator by following the instructions for use. After an adjustment is made, inspect the 
EasyStand to ensure all hardware and adjustable components are properly secured per the instructions 
for use. Replacement of worn parts, major adjustments, or any other important corrections should be 
handled by an authorized supplier or service center. Only Altimate Medical, Inc. approved replacement 
parts should be used to ensure safety and performance.

Do not use the EasyStand without consulting your physician . Failure to consult a qualified 
physician prior to using the EasyStand could lead to severe medical problems . We recommend 
that a good standing program be established and supervised by a qualified physician or physical 
therapist .

Join the Standing Community!
Our goal is simple - to empower people with disabilities to stand up for their health. Connect, share, and 
learn from other people with disabilities, special needs parents, clinical professionals, and the experts 
in standing on our social networking communities. By keeping your standing program “top of mind” you 
will have the greatest chance of maintaining a standing routine that maximizes all the benefits of 
standing! 

Read, share, and comment on the EasyStand Blog at blog.EasyStand.com 

Find EasyStand on these social networking sites:

Introduction

Page 1 - Introduction

Maintenance - Page 22

Cleaning/Disinfecting 

•   Clean the EasyStand Glider with a mild, non-abrasive disinfectant or soap and water.
•   Check the EasyStand at least weekly to ensure all nuts, bolts and adjustable parts are tightened securely, failure    
 

to do so may result in parts falling off which may cause a choking hazard for children.

•  Do not use bleach or any other bleach-based cleaners on the EasyStand. 
•  Disinfecting wipes are safe to use on metal and non-porous surfaces. 
•  Use a damp sponge with mild soap or a mild, non-abrasive disinfectant.   
•  Spot clean Neoprene with mild soap and water.  
•  Neoprene covers may be removed and hand washed with water. Air dry the covers. Always use care when  

 

 

removing and replacing the covers.  

•  Do not dry clean any upholstery items.
•  If the EasyStand is being used with one user, clean the EasyStand on a monthly basis or as needed
•  If the EasyStand is being used in a multiple user environment, the EasyStand should be cleaned after each use.

Urethane Coated Fabrics 

(Dartex or generic equivalent)

 

•  The recommended cleaning and disinfection instructions should be followed whenever possible to avoid damage    
 

to the PU membrane which could lead to premature failure of the cover.

•  For superficial dirt, use a disposable wipe and a warm solution of neutral detergent. Do not use abrasive cleaners. 
•  Disinfect using a 0.1% sodium hypochlorite solution (1000ppmm chlorine), stronger bleach solutions not 
 recommended. 
•  Thoroughly rinse any remaining cleaning solution with water and dry fabric before re-use or storage.
•  Do not iron.

Hydraulic Actuator Maintenance 

•   An initial break-in period of approximately 50-100 full cycles is normal for the hydraulic unit. The very close fit of    
 

its mating parts gives each pump its own break-in “feel” and signature sound, signaled by a squeaking or a 

 

less-than-smooth stroke.

•   If the pump is leaking oil, call your local EasyStand Supplier for a replacement. The serial number on the   

 

 

EasyStand will be needed.

•   Do not exert excessive force on the pump handle or use the pump handle to move an EasyStand. This may cause   
 

damage to the pump handle and/or the pump mechanism and will void any warranty.

•   User maintenance is not required or recommended on the hydraulic pump. Never apply lubricant, household oil,    
 

grease, or a similar product to any part of the unit. Some compounds may react with the seals used in the 

 

manufacture of this product. This may damage the unit and will void any warranty.

•   Do not exceed the rated weight limit.
•   If the EasyStand was shipped and exposed to a temperature of less than 50 degrees Fahrenheit, the hydraulic unit  
 

must warm to room temperature before use. Failure to follow this step could damage the unit and render it 

 inoperable.
•  Once the standing position is reached, the pump may settle back one stroke, this is normal.
•   It is not always necessary to push the pump handle all the way back to lower the seat. Try  pushing the handle  

 

 

back in different positions if it is difficult to lower the seat. 

•   If the unit is in the standing position and unoccupied, it is necessary to apply pressure to the seat of the unit while 
 

pushing the pump handle back.

Purging The Hydraulic Actuator

The hydraulic pump on the EasyStand has a high probability of getting air trapped in the internal workings of the unit. 
This is usually caused by in-transit handling and is quite normal. Please follow the simple steps below, as manual 
hydraulic systems may never operate properly unless the entrapped air is correctly purged. The unit must be at room 
temperature for this procedure.

1. Pump the handle until the full extension is reached.
2. Now continue pumping the handle slowly for an additional 10 strokes. This will properly purge the system.

Maintenance

Содержание Glider Medium

Страница 1: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Страница 2: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Страница 3: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Страница 4: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Страница 5: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Страница 6: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Страница 7: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Страница 8: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Страница 9: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Страница 10: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Страница 11: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Страница 12: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Страница 13: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Страница 14: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Страница 15: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Страница 16: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Страница 17: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Страница 18: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Страница 19: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Страница 20: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Страница 21: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Страница 22: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Страница 23: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Страница 24: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Страница 25: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Страница 26: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Страница 27: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Страница 28: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Отзывы: