CLEANING AND DISINFECTION
Before starting the cleaning and disinfection procedure, disconnect the
sensor from the USB port of the computer to avoid damage to your
equipment.
For your protection, during the cleaning and disinfection process, it´s
mandatory the use of IPEs such as disposable gloves and safety
glasses.
The cleaning and disinfection process must be performed at each patient change
and before the storage of the product.
When starting the process, check the existence of visible dirt, such as blood or
saliva.
Clean carefully the New IDA sensor and the first 30cm of the USB cable.
For the cleaning, use a clean and soft cloth damp with neutral soap and then
perform the drying with a clean and soft cloth or with a paper towel.
For the disinfection process of the New IDA sensor, use disinfectants towels that
have active components based on
Didecyldimethylammonium chloride
respecting the
contact time indicated by the manufacturer.
There's no cycle limits or time of application that the New IDA sensor and its parts
can tolerate during the cleaning and disinfection process, following the instructions of
this manual.
Don't use sanitizing or cleaning agents as a spray.
Don't dip the New IDA sensor in liquid sanitizing.
The New IDA sensor cannot be sterilized.