background image

12

Operating manual EE23 humidity / temperature sensor

For applications where condensation is likely to happen, certain ways of mounting  are required.

•  For probe hanging onto its cable from the ceiling, use the drip water protection HA010503 (not inclu-

ded in the scope of supply, see “spare parts / accessories”). This protects the probe and the sensing 

head against water dripping along the cable. 

•  For horizontal mounted probe, a drip water point should be made just before the probe.

13 

(0.51“)

85 

(3.3“)

34 

(1.3“)

horizontal

hanging onto the probe cable

stainless steel

mounting flange

drip water loop

drip water

protection

Fig. 9 

Mounting the sensing probe

4.3  Electrical Connection

Power supply and outputs

Alarm output

output and supply

alarm output

Relais status for 

RH < threshold

Power supply

Fig. 10 

Connection diagram

4.4  Connection Versions

standard

option E4

option AM3

1) Two M16x1.5 cable glands (one of them mounted onto the enclosure)

2) Power analogue outputs: cable connector, 5-poles, straight M12 (Lumberg RKC 5/7)

3) Analogue output: cable connector, 5-poles, straight M12 (Lumberg RKC 5/7)

4) AC power supply: cable connector, 3-poles, straight (7/8ʺ-16UN)

Содержание EE23 series

Страница 1: ...BA_EE23_B2_e v1 0 technical data are subject to change 194768 Operation Manual EE23 Humidity Temperature Sensor...

Страница 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Страница 3: ...4 Intallation 10 4 1 Mounting the Enclosure 10 4 2 Mounting the Sensing Probe 11 4 3 Electrical Connection 12 4 4 Connection Versions 12 5 Calibration Adjustment 14 5 1 Definitions 14 5 2 Adjustment...

Страница 4: ...ons The instructions shall be observed in order to reach optimal performance of the device 1 2 Safety Instructions 1 2 1 General Safety Instructions Avoid any unnecessary mechanical stress and inappro...

Страница 5: ...mponents of the device must be separated according to local recycling regulations The electronics shall be disposed of correctly as electronics waste 2 Scope of Supply Included in all versions Accordi...

Страница 6: ...the optional display 3 Status LEDs D1 red continuously on during the calibration routine one short flash confirms reset to factory calibration D2 green flashes during normal operation continuously on...

Страница 7: ...es that the device in calibration adjustment mode 2 indicates measured value above the alarm threshold with alarm module only visual feedback when depressing the S1 push button Fig 1 indicates measure...

Страница 8: ...e optional display 2 Status LEDs Green LED flashing Supply voltage applied Microprocessor is active Red LED constantly lit Error category 1 non critical error can be solved by the user flashing Error...

Страница 9: ...cription Error code display Error category Recommended action Short circuit on voltage output 11 ERROR 01 1 Check wiring of outputs Short circuit on voltage output 21 Short circuit on both voltage out...

Страница 10: ...s pointing downwards or horizontal 3 Wire the terminals inside the back cover according to Fig 10 Connection diagram on Page 12 4 Insert the middle active part into the back cover By this the connecti...

Страница 11: ...mportance for accurate mea surement to avoid T gradients along the probe In case of large T difference between the two sides of the wall it is highly recommended to insert the probe completely up to t...

Страница 12: ...e 13 0 51 85 3 3 34 1 3 horizontal hanging onto the probe cable stainless steel mounting flange drip water loop drip water protection Fig 9 Mounting the sensing probe 4 3 Electrical Connection Power s...

Страница 13: ...2 OUT1 1 plug for 100 240 V metal enclosure front view Description Pins grounding PE 1 phase L1 2 neutral wire N 3 plug for 100 240 V polycarbonate enclosure front view Description Pins phase L1 1 neu...

Страница 14: ...54 F The 1 point adjustment should be used only when the RH and or T range of interest is rather narrow The RH resp T adjustment point should be ideally in the middle of the range of interest Example...

Страница 15: ...the jumper to CAL T 2 Allow for the probe to stabilize at the desired RH T for min 30 minutes 3 For adjustment point 50 RH in the upper half of the T output scale press BUTTON S1 for min 3 seconds to...

Страница 16: ...eeded with an E E original one A polluted filter cap causes longer response time of the device For cleaning of the sensing head please see Cleaning Instructions at www epluse com ee23 7 Troubleshootin...

Страница 17: ...T6 D03M EE23 for T5 D05M EE23T5 Display housing cover in polycarbonate for T1 T2 T4 and T6 D03P EE23 for T5 D05P EE23T5 Mounting flange for probe diameter 12 mm HA010201 Mounting flange 5mm for model...

Страница 18: ...tage for 0 1 V 0 5 V outputs 10 5 35V DC or 12 28V AC for 0 10 V 0 20 mA 4 20 mA outputs 15 0 35V DC or 15 28V AC 100 240V AC 50 60Hz supply module optional Current consumption for voltage output for...

Страница 19: ...ing Range Humidity Sensor Relative humidity RH Temperature degC The gray area shows the allowed measurement range for the humidity sensor Operation beyond this range does not destroy the sensing eleme...

Страница 20: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Отзывы: