Operating manual EE23 humidity / temperature sensor
5.3 Adjustment of the EE23 Model T5 via the USB Service Interface
1.
Download and install the EE-PCS Product Configuration Software from www.epluse.com/configurator
2.
Connect the USB service interface of EE23-T5 to the PC.
3.
Start the EE-PCS software.
4.
Select the desired adjustment mode and follow the instructions of the EE-PCS.
6 Maintenance
When employed in dusty, polluted environment:
•
The filter cap shall be replaced as needed with an E+E original one. A polluted filter cap causes
longer response time of the device.
•
For cleaning of the sensing head please see “Cleaning Instructions” at www.epluse.com/ee23.
7 Troubleshooting
See status LEDs at „3 Operating Components“ on Page
6 and error messages on the display at „3.4
Display for Model T5“ on Page
9.
7.1 Fuse Replacement for Option AM3
If the green LED on the PCB is not flashing with the supply voltage switched on check the fuse and
replace if required.
Fuse secondary:
250mA / T UL248-14
Recommended replacement types:
Series: MSTU 250 Manufacturer: Schurter
Order No.: 0034.7109
Series: 374
Manufacturer: Littelfuse
Order No.: 374 0250
Fig. 11
Fuse replacement