CAUTION:
Check correct operating voltage on the side label before
connecting power. A UL listed 1A anti-surge fuse, rated 250V (for AC)
65V (for DC) should be fitted to the power input. An IEC60947-1 &
IEC60947-3 compliant isolation switch should be fitted close to the unit,
in easy reach of the operator, and appropriately marked.
All connections to the device must be made through a spade format or
similar connection, with connection to the spade terminal touching both the
insulation and conductor material.(Use a standard crimping tool) All
connections must be Mechanically secured so as to prevent any wiring
becoming loose and coming in contact with other wires or the instrument
casing.
The above applies to any and all connection to hazardous mains supply
either direct or indirect (Through a switch (Relay))
Main Terminals
CAUTION:
External computing devices connected to the
communications port must comply with the standard, UL 60950.