Page 76 of 146
198-1600-00 Rev F
28.
From the “Option Card Calibration Menu” rotate the current encoder until the display
shows “CAL” “CL15”. Press the voltage encoder switch to view and change the lower
voltage control analog input calibration point. The top display will show “E.Lo CtL
CL15”, the bottom display shows the calibration point value.
29. Set the calibrator output to 10% of the analog input range (5.6mA for 4-20mA or 0.5V
for 0-5V or 1V for 0-10V).
30. Put the supply in REMOTE mode (press the RMT/LCL switch so that the RMT/LCL
lamp is on) to begin calibrating the lower voltage control analog input calibration point.
31. The lower voltage control analog input calibration value (CL15) is automatically
updated to reflect the connected calibrator output. Wait a few seconds for the
calibration value to stabilize and then press the voltage encoder switch to accept the
new value and return to t
he “Option Card Calibration Menu”.
32.
From the “Option Card Calibration Menu” rotate the current encoder until the display
shows “CAL” “CL16”. Press the voltage encoder switch to view and change the lower
voltage read back analog output calibration point.
The top display will show “E.Lo rEAd
CL16”, the bottom display shows the calibration point value.
33. Put the supply in REMOTE mode (press the RMT/LCL switch so that the RMT/LCL
lamp is on) to begin calibrating the lower voltage read back analog output calibration
point.
34. Adjust the current encoder until the calibrator input reads the proper value for 10% of
the analog output control signal (5.6mA for 4-20mA or 0.5V for 0-5V or 1V for 0-10V).
35. Press the voltage encoder switch to accept the new calibration value and return to the
“Option Card Calibration Menu”.
36.
From the “Option Card Calibration Menu” rotate the current encoder until the display
shows “CAL” “End”. Press the voltage encoder switch to exit the “Option Card
Calibration Menu”.
37. If changes were made to any calibration values you will be prompted to store the
changes. Select “yES” to accept the changes or “no” to cancel the changes to the
calibration values.