NOTE:
When the the output type is set to “custom” the current and voltage settings can not be
changed using the front panel display; the current and voltage settings are defined by the selected
waveform profile.
6.4.6. Ramp Settings
The power supply can be configured to ramp current or voltage from zero to the active setting
when the power supply starts a cycle. The definition of the start of a cycle is dependant on the
cycle control mode. For manual cycles the start of a cycle is any time the power supply changes
from standby to operate mode. For timed cycles (RTC or ATC) the start of a cycle is when the
power supply changes from a non-paused standby mode to operate mode (a new cycle is
started). If the output type is set to “Pulse” or “Custom Waveform” the amplitude of the waveform
is ramped (the shape of the waveform is maintained throughout the ramp). Ramping is
accomplished by ramping the current and/or voltage settings at the specified ramp rate, as such
regulation modes are still valid as defined by
6.8 Setting Adjustment and Regulation Modes
Use the “Ramp Setup Menu” in the “Unit Setup Menu” to configure ramp settings. The following
procedure describes how to change ramp settings. See